2 storage and handling – Bio-Rad ProteinChip Qualification and Calibration Kits User Manual
Page 6
Table 1. Summary of procedures and tests (
continued).
R
Reecco
om
mm
meen
nd
deed
d
C
Co
on
nssu
um
maab
bllee
E
Essttiim
maatteed
d T
Tiim
mee
T
Teesstt
FFrreeq
qu
ueen
nccyy
U
Usseed
d
R
Reeq
qu
uiirreed
d
D
Deessccrriip
pttiio
on
n
T
Teesstt 2
2
Mass drift and
Weekly
ProteinChip
20 min to 1 hr
This test uses the ProteinChip
resolution at
peptide
peptide standard array to
5.96 kD
standard array
measure the mass drift and
resolution of insulin. These
measurements are compared to a
specification, and a pass/fail
disposition is obtained.
T
Teesstt 3
3
Resolution at
Weekly
ProteinChip
20 min to 1 hr
This test uses the ProteinChip
1 kD
peptide
peptide standard array to
standard array
measure the resolution of
Arg-vasopressin. These
measurements are compared to
a specification, and a pass/fail
disposition is obtained. The test
is run in higher-resolution
(lower source voltage) mode.
T
Teesstt 4
4
Mass
Weekly
ProteinChip
30 min to 1 hr
This test evaluates the mass
accuracy
peptide
accuracy when compared to
standard
internal and external
array
calibrations. These measurements
are compared to a specification,
and a pass/fail disposition is
obtained.
1.2 Storage and Handling
Store the arrays supplied in the kit in the original packaging at room temperature and in a dry
and dark location. Each array is individually wrapped. Do not open the array packaging until
ready to use. Take care when handling the ProteinChip arrays to avoid contaminating or disturbing
the matrix crystals on the surface of the arrays.
3
10010680A.qxp 8/17/2007 1:58 PM Page 3