Bio-Rad ProteinChip Qualification and Calibration Kits User Manual
Page 29
9. Zoom in and check that all 7 peaks in all 8 spectra are marked correctly. Failure to do so
may result in incorrect results. Close the plot.
Mass Accuracy — External Calibration Test
Perform this test BEFORE the internal calibration test.
1. Click on the spectrum for spot D.
2. Select Spectra > Calibrate > Apply Calibration Protocol.
3. Select OQ Mass Cal.
4. Click OK. The spectrum flag changes from yellow to green.
5. In the main window, make sure that only the spectrum from spot D is selected. Click the
Copy Calibration button. The spectrum flag on spot D reverts to yellow.
6. Select all 8 spectra and click the Apply Calibration button. (Do not select the first
optimization spectrum.)
7. Export the data following the instructions in Appendix C and selecting the parameters for
export in the following order:
•
Array bar code
•
Spot name
•
Substance mass
•
Intensity
•
Resolution
•
S/N
•
Peak #
8. Paste the data into the ProteinChip SELDI OQ form, data sheet “Test 4 Mass Accuracy”.
Paste the data into cell M41.
26
Spectrum
flag
10010680A.qxp 8/17/2007 1:59 PM Page 26
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- iMark Microplate Absorbance Reader (112 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- Model EP-1 Econo Pump (23 pages)
- Model EP-1 Econo Pump (1 page)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)