Overview, Obtain static ip address, Assign static ip to the reader – Bio-Rad ProteinChip Data Manager Software User Manual
Page 43: Assign static ip to the reader 38

Page 38
Networking the ProteinChip SELDI System
ProteinChip SELDI System: Reader Guide
2.
Enter the hostname of the server into the Data Manager Server hostname
field.
3.
Click the Update button to save the changes.
NOTE
If the server hostname does not save properly (i.e. reverts back to its original
settings, please see Instrument Options in the Troubleshooting section,
starting on page 48.)
Networking
the
ProteinChip
SELDI System
Using Static
IP Addresses
Overview
1.
Obtain static IP addresses from the local IT group for server and the reader
2.
Assign static IP to reader
3.
Assign static IP to server computer
4.
Redirect client computers to new server address
5.
Obtain new license file (if necessary)
6.
Connect to the reader
7.
Edit the Instrument Options settings
Obtain Static IP Address
Static IP addresses are required for the reader and server in order to ensure
uninterrupted network communication between the ProteinChip SELDI system
reader and the server.
You will need the following pieces of information for the server and the reader:
•
IP address
•
Subnet mask
•
Default gateway
•
DNS domain name
•
DNS server IP address
Assign Static IP to the Reader
The reader should already be configured with the IP address 192.168.1.200
when it arrived, and should already be set up to communicate with the
supplied workstation. The workstation will be used to update the reader
settings.