Bio-Rad ProteinChip Data Manager Software User Manual
Page 27

Page 22
ProteinChip SELDI System: Reader Guide
10. Replace the load lock door by screwing it clockwise back on to the post.
11. Tighten the door until it seals against the O-ring.
12. Unscrew the load lock door about 1/2 turn until it is in the open position
(cutout in door should match opening in load lock).
13. Carefully replace the pawl drive assembly, lining up the posts on the load
lock door with the pawl drive plate. Take care not to catch the load lock
door home sensor wires between the pawl drive plate and the autoloader
mounting plate (Fig. 21).
Fig. 21: Load lock door sensor wires.
See also other documents in the category Bio-Rad Measuring instruments:
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- iMark Microplate Absorbance Reader (112 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)