beautypg.com

Bio-Rad AG® MP-1M Anion Exchange Resins User Manual

Page 7

background image

batch as well as column applications. Medium mesh is
an ideal, general purpose particle size for use in analyti-
cal and preparative scale column chromatography. Fine
mesh material (200-400 and minus 400 mesh) is used
for high resolution analytical separations.

9

The AG 1 and AG MP-1 resins are available in sev-

eral particle size ranges. The flow rate in a chromato-
graphic column increases with increasing particle size.
However, the attainable resolution increases with
decreasing particle size and narrower size distribution
ranges. Particle size is given either in mesh size or
micron size. The larger the mesh size number, the small-
er the particle size. Table 4 shows wet mesh and equiva-
lent micron diameters.

Table 4. Wet Mesh and Equivalent
Micron Diameters

Wet Mesh

(U.S. Standard)

16

20

40

50

80

100

140

200

270

325

400

Micron Diameter

(1 µm = 0.001 mm)

1,180

850

425

300

180

150

106

75

53

45

38

Large mesh material (20-50 and 50-100 mesh) is

used primarily for preparative applications and batch
operations where the resin and sample are slurried
together. Medium mesh resin (100-200) may be used in

8

LIT212C 6/17/98 12:25 PM Page 8