5.2 Column Method

The column method involves pouring a column with

the resin and passing the sample through to achieve the
separation. Particle size will determine the flow rate,
which will affect the separation. The resin should be in
the correct ionic form and equilibrated prior to adding
the sample.

1. Calculate the amount of resin required based on the

expected resin capacity and sample concentration. If
the sample ionic concentration is unknown, begin
with 5 grams of resin for 100 ml of sample, and then
optimize the volumes after looking at the results of
the first separation.

2. Insure that the resin is in the proper ionic form to

allow the sample ions to be exchanged onto the
resin. If conversion of the resin into another ionic
form is necessary, use the guidelines described for
resin conversion (see Table 5).

3. Prepare the initial buffer so that the pH and ionic

concentration will allow the sample ions to be

15

Section 5
Instructions for Use

AG 1 and AG MP-1 resins may be used in a batch

method or a column method. The batch method consists
of adding the resin directly to the sample and stirring.
The column method requires packing a column with
resin, and passing the sample through.

5.1 Batch Method

The batch method is performed by adding the resin

directly into the sample and stirring. The resin should be
in the correct ionic form prior to beginning.

1. Weigh out about 5 grams of resin for every 100 ml

of sample. For larger scale applications or when an
exact amount of resin is needed, calculate the resin
volume based on the resin capacity.

2. Add resin to the sample and stir or shake gently for

1 hour.

3. Filter or decant the sample from the resin.