Fig. 2. connecting to an mv-6 valve – Bio-Rad Profinity eXact™ Purification and Tag Cleavage Consumables User Manual
Page 7

Fig. 2. Connecting to an MV-6 valve.
3. Connect the inlet of the cartridge to the male
luer fitting on the MV-6 sample inject valve (see
Figure 2). If not using the MV-6 sample inject
valve, connect a barb to male luer fitting on the
1.6 mm ID tubing, then connect to the top of
the female luer on the Bio-Scale Mini cartridge.
For optimum performance, a cartridge should
be mounted vertically with the arrow on the
cartridge pointing downward (see Figure 3).
TO
COLUMN
TO
BIOLOGIC LP
SYSTEM OR ECONO
GRADIENT PUMP
TO
BIOLOGIC LP
SYSTEM OR ECONO
GRADIENT PUMP
MV-6
INJECT
VALVE
SAMPLE LOOP
WASTE
INJECT
PORT
FILL
TO
COLUMN
MV-6
INJECT
VALVE
SAMPLE LOOP
WASTE
INJECT
PORT
FILL
"INJECT" POSITION
"FILL" POSITION
9
Fig. 1. BioLogic LP system setup. (Use orange lock rings
and medium size barb fittings with 1.6 mm tubing.)
2.
To maximize gradient accuracy and to apply
samples efficiently, install 1.6 mm ID tubing
from the pump to the MV-6 sample inject valve
(if available). If using the MV-6 sample inject
valve, turn the knob counterclockwise as far as
it will go so it corresponds to the printed diagram
on the valve (see Figure 2).
1
Ala
rm
2
3
4
5
6
7
8
0
9
C
.
Platen pressure
screw
Lock-ring
Tubing
Luer fittimg
See detail
8
10011164A.qxp 11/19/2007 9:54 AM Page 8