Purification Protocol

1. Equilibrate the cartridge with 10 column volumes

(CV) of bind/wash buffer at 3 ml/min (10 ml/min).

2. Load the sample lysate at 1 ml/min (5 ml/min).

3. Wash the cartridge with 10 CV of bind/wash

buffer at 3 ml/min (10 ml/min).

4. Equilibrate cartridge with 2 CV of elution buffer

at 3 ml/min (10 ml/min)*.

5. Stop flow and incubate for 30 minutes

**

.

6. Elute purified protein with 5 CV of elution buffer

at 3 ml/min (10 ml/min).

* Alternative elution protocol: Pump 5 CV of elution buffer at

0.1 ml/min (0.5 ml/min) at room temperature.

**Increased incubation time, 2–4 hr at 25°C or 12–24 hr at 4°C,
may be necessary for complete cleavage of some proteins.

21

5.

Remove the supernatant, and filter through a
0.45 µM filter before applying to the cartridge.

Section 8
Cartridge Preparation and
Purification Protocol

To prepare a cartridge, remove the end plugs and
connect the cartridge to the chromatography system. The
cartridge is now ready for use. The following 1 ml
cartridge purification protocol is a general guideline
for first time users. Flow rates for 5 ml cartridges are
shown in parentheses. The kinetics of cleavage are
target protein specific and may require optimization
to maximize yield. See Figure 6 for a representative
purification profile using the following protocol. Refer
to the Profinity eXact System Manual for complete
details.