beautypg.com

Table 2. profinity exact resin specifications – Bio-Rad Profinity eXact™ Purification and Tag Cleavage Consumables User Manual

Page 5

background image

Table 3. Buffer and chemical compatibilities for
Profinity eXact cartridges.

Reagent
Group

Description

Stability

Triggering Ions

F-, Cl-, N

3

-, NO

2

-, CO

2

-

Do not use in lysis and
wash buffers.
See Table 5 for use as
trigger in elution buffer.

Salts

Sodium Acetate

£3M

NaCl or KCl

Do not use in lysis and
wash buffers.

Buffers

Tris-HCl

Substitute Tris-acetate
or Tris-phosphate.

Acids

HCl

Do not use. Substitute
acetic or phosphoric
acid.

Detergents

Non-ionic

£5% (w/v)

Zwitterionic

£5% (w/v)

Ionic

Do not use.

Protease

PMSF, CalBioChem cocktail,

inhibitors

Roche Protease Inhibitor

1X

cOmplete mini tablet

BD Biosciences cocktail

2X

Lysis Solutions

Lysis & Extractrion Reagent

1X

(Bio-Rad), B-PER phosphate
(Pierce), BugBuster (EMD),
FastBreak Cell Lysis (Promega)

ReadyPreps Lysis (Epicentre),

Do not use.

CelLytic Express (Sigma)

5

Table 2. Profinity eXact resin specifications.

Functional ligand

Mutant subtilisin

Base bead

Superflow 6% agarose

Particle size range

60–160 µm

Recommended linear flow rate

<1,000 cm/hr at 25°C

Maximum operating pressure

45 psi

Chemical compatibility

See Table 3

Storage

4°C

Shelf life in storage buffer

>1 year at 4°C

Operational temperature

4–40°C

4