5.1 With a Liquid Chromatography System

1. Set the flow rate to the desired flow.

a.

10 ml cartridge: 0.5–4.0 ml/min at room
temperature, 0.5–2.0 ml/min at cold room
temperature.

b.

50 ml cartridge: 2.5–20 ml/min at room
temperature, 2.5–10 ml/min at cold room
temperature.

2. Wash the prepared cartridge with degassed

buffer as indicated below.
a.

10 ml cartridge: 10 ml

b.

50 ml cartridge: 50 ml

3. Inject up to the amount of sample indicated

below, discarding the first amount of effluent
indicated below.
a.

10 ml cartridge: 6 ml of sample, discarding
the first 3 ml of effluent

b.

50 ml cartridge: 30 ml of sample, discarding
first 15 ml of effluent

4. Monitor the cartridge effluent with a UV monitor

or conductivity monitor.

15

10006995B.qxp 6/13/2007 9:00 AM Page 15