3.3 FPLC Systems
The Luer to M6 adaptor fittings kit (catalog
#732-0111) provides fittings necessary to connect
the cartridge to the M6 fittings found on FPLC or
related systems.
Alternatively, connection can be made by using two
Upchurch P-621, ¼–28 to metric adaptors, one
Upchurch P-619, ¼–28 to male Luer, and one
Upchurch P-628, ¼–28 to female Luer. Assemble
the Luer fitting to the ¼–28 metric adaptors. Attach
the adaptor with the male Luer to the column inlet
line of the FPLC system and the adaptor with the
female luer to the FPLC column outlet. To prevent
tubing or cartridge failure, do not exceed the
maximum recommended flow rate.