Section 12

Troubleshooting Guide

This troubleshooting guide addresses problems that may be encountered
during a protein coupling reaction. If you experience any of the problems listed
below, review the possible causes and solutions provided. Contact Luminex
Tech Support for technical issues related to coupling of oligonucleotides
(source: http://www.luminexcorp.com/support/tech_tips.html).

Possible Causes

Possible Solutions

Low MFI signal in validation procedure

EDAC and S-NHS may have
expired

Purified protein was not used
immediately during the conjugation
procedure or was not kept on ice

S-NHS was not added to the
beads immediately after adding
EDAC

Wrong antibody was used in the
validation procedure

Use fresh EDAC and S-NHS for
each conjugation

Keep the protein on ice during the
conjugation procedure

Add S-NHS to the beads
immediately after adding EDAC

Ensure that the correct species-
specific antibody is being used for
validation (if an antibody is coupled)