Bio-Rad Bio-Plex® Precision Pro™ Human Cytokine Assays User Manual

Page 12

background image

10

Section 8

Assay Instructions

The following instructions apply to Bio-Plex

®

Precision Pro™ cytokine

assays. All of the necessary components are provided premixed for ease
of use.

Plan Experiment
1. Assign which wells of a 96-well plate will be used for each standard,

control, and sample (see the example below).

2. Determine the total number of wells that will be used in the assay.

Include a 25% excess (or add 2 wells for every 8 wells used) to
ensure that enough diluted coupled beads, detection antibodies,
and streptavidin-PE are prepared.

Example Plate