beautypg.com

Bio-Rad Human MMP and TIMP Assays User Manual

Page 35

background image

33

Possible Causes

High Background Signal

Incorrect buffer was used
(for example, assay buffer
used to dilute standards)

Accidentally spiked blank
wells

Detection antibodies or
streptavidin-PE incubated
too long

Poor Recovery

Expired Bio-Plex reagents
were used

Incorrect amounts of
components were added

Microplate shaker set to an
incorrect speed

High end saturation of the
standard curve

Quality controls do not fall
within expected ranges

Possible Solutions

Use standard diluent HB or diluent similar
to final sample matrix to dilute standards.

Do not add any antigens to the
blank wells.

Follow the procedure incubation
time precisely.

Check that reagents have not expired. Use
new or nonexpired components.

Check your calculations and be careful to
add the correct volumes.

Check the microplate shaker speed and
use the recommended setting. Setting the
speed too high may cause splashing and
contamination. Use the recommended
plate shaker. Setting the speed too low may
cause low assay signal and false plateau or
high end saturation of standard curves.

Make sure that correct shaker speed and
incubation times are used. Remove S1
from data analysis if needed.

Make sure that the controls are
reconstituted at the same time as standards
and in the same diuent (standard diluent
HB). Incubate for precisely 30 min.