Bio-Rad PowerPac™ HV High-Voltage Power Supply User Manual
Page 24

Step Procedure
Description
5.
Method steps
Press EDIT to enter the edit loop.
New method
• Press the up and down arrows to select an
existing step or select the (EMPTY) step to add a
new step
• Press EDIT again to enter the edit loop for the
selected step
Existing method
6.
Set the constant
Use the soft keys to select the constant parameter
parameter for the
and the keypad to enter its value. Note that the
current step
non-constant values are listed on the right side of
the screen and the time mode at the bottom.
The S#/# indicates which step is being modified. In
this case, S1/1 means editing step 1 of 1.
CONSTV:
Operates with constant voltage.
CONSTA:
Operates with constant current.
CONSTW: Operates with constant power.
Press EDIT to continue.
7.
Enter limits for the
Use the soft keys to select a parameter and the
non-constant
alphanumerical keypad to enter its value.
parameters for the
current step
The soft keys CONST_ or LIMIT_ change according
to the constant parameter selected in the EDIT
CONSTANT screen.
Note: The constant parameter value can be
changed at this point. Press its soft key and enter a
new value.
LIMITV
Change the voltage limit (or constant)
(CONSTV): value.
LIMITA
Change the current limit (or constant)
(CONSTA): value.
LIMITW
Change the power limit (or constant)
(CONSTW): value.
Press EDIT to continue.
16
METHOD: NAME
S1 (EMPTY)
*
METHODS
METHOD: NAME
S1 100V 00:30 HRS
S2 (EMPTY)
*
METHODS
DELETE
SAVE
500 mA
400 W
S1/1 EDIT LIMITS
TIME: UNTIMED
CONSTV LIMIT
A
LIMITW
PowerPac HV – Instruction Manual