Troubleshooting – Bio-Rad 2-D Electrophoresis Workflow User Manual
Page 22
![background image](/manuals/600212/22/background.png)
18
Section 4
Troubleshooting
continued
Problem
Possible Cause
Remedy
Bands or spots No protein on gel
Verify that there is actually
not visible
protein on the gel by staining
with another method such as
Bio-Safe
™
Coomassie stain.
Malfunctioning
Check instrument manual
imaging system
for troubleshooting, or
contact the imaging
instrument manufacturer.
Poor staining
Insufficient
Stain sensitivity maximizes
sensitivity
staining time
after 90 min.
Dirty staining
Make sure that the staining
trays
trays and other equipment
have been thoroughly
cleaned with laboratory
glassware cleaner.
Insufficient
Follow the recommendations
stain volume
for stain volume appropriate
to the gel size (Section 2.2).