Bio-Rad Silver Stains User Manual
Page 7
Working Solution Volumes
For best results, prepare each working solution no more than 24 hr before use. Add the image developer
concentrate to the developing working solution within 15 minutes of use. Note: The volume corresponding
to water (Table 2, 3 and 4) indicates the final volume of the working solution.
Please refer to the appropriate section below to determine the proper working solution volumes for staining
1) 12 PROTEAN gels in a Dodeca stainer, 2) fewer than 12 gels in a Dodeca stainer, or 3) individual gels in a
user-provided staining container.
1)
Dodeca stainer with 12 trays. When using the full capacity of the Dodeca stainer, prepare each
working solution as directed by Table 2. Please note that the cover tray is not counted.
Table 2. Dodeca stainer solution volumes.
Working Solutions
Reagents
Small Dodeca Stainer
Large Dodeca Stainer
12 Trays
12 Trays
Fixing
Ethanol
2.8 L
4.0 L
Glacial acetic acid
700 ml
1.0 L
Water
To 7 L
To 10 L
Sensitizing
SC (10×)
700 ml
1.0 L
BRC (100×)
70 ml
100 ml
Ethanol
2.1 L
3.0 L
Water
To 7 L
To 10 L
Staining
SRC (50×)
140 ml
200 ml
Water
To 7 L
To 10 L
Developing
DBC (10×)
700 ml
1.0 L
IDC (5,000×)
1.4 ml
2.0 ml
BRC (20,000×)
350 µl
500 µl
Water
To 7 L
To 10 L
Stopping
Glacial acetic acid
350 ml
500 ml
Water
To 7 L
To 10 L
5