3)
Individual gels. Prepare working solutions according to Table 4. To stain multiple gels, multiply the
solution volumes by the appropriate number of gels. The working solution volumes listed below are
recommended volumes — the actual required volume may vary depending on the dimensions of the
staining container used. For all cases, sufficient volumes of working solution should be prepared to
completely submerge the gel to be stained.
Table 4. Suggested working solution volumes for individual gels.
Working Solutions
Reagents
Individual Mini or
Individual Large Gels
Criterion
™
Gels
16 x 16 cm or larger
(250 ml)
(1 L)
Fixing
Ethanol
100 ml
400 ml
Glacial acetic acid
25 ml
100 ml
Water
To 250 ml
To 1 L
Sensitizing
SC (10×)
25 ml
100 ml
BRC (100×)
2.5 ml
10.0 ml
Ethanol
75 ml
300 ml
Water
To 250 ml
To 1 L
Staining
SRC (50×)
5 ml
20 ml
Water
To 250 ml
To 1 L
Developing
DBC (10×)
25 ml
100 ml
IDC (5,000×)
50.0 µl
200.0 µl
BRC (20,000×)
12.5 µl
50.0 µl
Water
To 250 ml
To 1 L
Stopping
Glacial acetic acid
12.5 ml
50.0 ml
Water
To 250 ml
To 1 L
8
4110150B.qxd 7/31/2003 1:39 PM Page 8
- TransFectin™ Lipid Reagent (2 pages)
- Gene Pulser MXcell™ Electroporation System (66 pages)
- Gene Pulser MXcell™ Electroporation System (19 pages)
- Gene Pulser Xcell™ Electroporation Systems (83 pages)
- Gene Pulser® Electroporation Buffer (2 pages)
- MicroPulser™ Electroporator (31 pages)
- Helios® Gene Gun System (52 pages)
- PDS-1000 / He™ and Hepta™ Systems (51 pages)
- TGX™ FastCast™ Acrylamide Solutions (2 pages)
- Criterion™ TGX Stain-Free™ Precast Gels (16 pages)
- Criterion Stain Free™ Tris-HCl Gels (96 pages)
- Image Lab™ Software (212 pages)
- Gel Doc™ EZ System (22 pages)
- Mini-PROTEAN® TGX™ Precast Gels (52 pages)
- ChemiDoc™ XRS+ System (4 pages)
- ChemiDoc™ XRS+ System (50 pages)
- GS-800™ Calibrated Densitometer (444 pages)
- ChemiDoc™ MP System (8 pages)
- Image Lab™ Software (236 pages)
- ChemiDoc™ XRS+ System (42 pages)
- Criterion™ TGX™ Precast Gels (60 pages)
- Criterion™ Cell (13 pages)
- Image Lab™ Software (260 pages)
- Criterion™ XT Bis-Tris Precast Gels (26 pages)
- 2-D Electrophoresis Workflow (14 pages)
- 2-D Electrophoresis Workflow (20 pages)
- PROTEAN® i12™ IEF System (4 pages)
- ReadyPrep™ 2-D Starter Kit (28 pages)
- EXQuest Spot Cutter (81 pages)
- Fluorescent Protein Stains (27 pages)
- 2-D Electrophoresis Workflow (22 pages)
- Ready Gel® Zymogram Precast Gels (46 pages)
- Mini-PROTEAN 2-D Electrophoresis Cell (44 pages)
- Precision Plus Protein™ Prestained Standards (3 pages)
- Precision Plus Protein™ Unstained Standards (16 pages)
- Prestained SDS-PAGE Standards (3 pages)
- Silver Stains (20 pages)
- Unstained SDS-PAGE Standards (3 pages)
- Biotinylated Standards (3 pages)
- Biotinylated Standards (11 pages)
- IEF and 2-D Standards (3 pages)
- Mini-PROTEAN 2-D Electrophoresis Cell (20 pages)
- Mini-PROTEAN® Tetra Handcast Systems (10 pages)
- Mini-PROTEAN 3 Multi-Casting Chamber (10 pages)
- PROTEAN® Plus Hinged Spacer Plates and Combs (21 pages)