Troubleshooting – Bio-Rad S3™ Cell Sorter User Manual
Page 31
Instruction
Manual | 21
Troubleshooting
5
Table 7 describes possible errors and causes with troubleshooting steps.
Table 7. Troubleshooting table.
Error
Possible Causes
Troubleshooting Steps
Biosafety system icon
System not configured to have
Contact technical support to correct the
does not show up in
a biosafety system
configuration of the system
ProSort
™
software
There is no fan noise
System is very quiet
Stand closer to fan unit and listen for fan
coming from the unit
noise. There should be a quiet hum
Not plugged in correctly
Check AC power cord for a tight connection
on system and in the wall
HEPA may be full and system
Check the biosafety system status in ProSort
has shut down.
software to see if the system’s HEPA is at
less than 10% capacity. HEPA filter may need
replacement
There is a rattling noise
Loose fitting or latch
Check all connections and latches. The
coming from the air
system should have a quiet hum and low
generator fan unit
vibration
There is a large gap
Misaligned magnetic attachment
Re-attach the access door. Adjust by gently
between magnetic
bending the access door to eliminate the
connections on
gap. If the gap remains and the access door
access door
border matches with the panel borders, then
system is functional
The panels do not stay
Panels do not lie well against air
Roll the panels slightly tighter so they can lie
up when opened
generator and plastic top panel
evenly on the plastic top panel
The access doors are
S3 or S3e cell sorter is in the
Adjust the S3 or S3e cell sorter carefully to
not aligned with the
improper location
line up with the access doors as much as
system components
Improper location may not maintain safety