S3 biosafety system class i certification form – Bio-Rad S3™ Cell Sorter User Manual
Page 26
Service and Certification
16 | S3 Biosafety System
S3 Biosafety System Class I Certification Form
S3 Biosafety System Class I Certification Form
Customer Name
Phone
Company
Location
Enclosure SN
Inflow Velocity
Airflow readings, FPM __________________________
__________________________
__________________________
__________________________
__________________________
__________________________
__________________________
__________________________
__________________________
__________________________
Calculated average airflow
_________________________ FPM
Average airflow >135 FPM?
______ PASS or ______ FAIL
Measured area
______ ft
2
Area of perimeter base slots
______ 0.15 _________ ft
2
Total area (measured area + perimeter slots)
______ ft
2
Exhaust volume (average airflow x total area)
_________________________ CFM
Leak Integrity
Upstream concentration
______________________________
Upstream concentration >10 µg/L?
______ PASS or ______ FAIL
Exhaust discharge concentration
______________________________
Exhaust discharge <0.01% upstream concentration?
______ PASS or ______ FAIL
Smoke Containment
Smoke drawn into enclosure?
______ PASS or ______ FAIL
Certifier (Printed Name)
Certifier (Signature)
Date of Certification