Reference, Sample the following locations for 10 min each, Near sample station – Bio-Rad S3™ Cell Sorter User Manual
Page 29: Front, below bulk fluids door, On top, between nozzle and filter door, Right side, below vents, Left side, below vents
Contaminant Testing
Instruction
Manual | 19
Sample the following locations for 10 min each:
■
n
Near sample station
■
n
Front, below bulk fluids door
■
n
On top, between nozzle and filter door
■
n
Right side, below vents
■
n
Left side, below vents
■
n
Rear, next to the AC entry
After all samples have been collected:
24. Roll down both left and right panels.
25. Roll down both the front and the rear panels.
26. Run 10% bleach through the sample line for 10 min.
27. Run deionized water through the sample line for 10 min.
28. After 10 min of running DI water, ensure the event rate is under 5 eps.
29. Read each slide using a 10x objective.
30. Record the number of Glo Germ particles on a 1 x 1” area on the microscope slide using
the containment testing sheet.
Results are informational and not pass/fail.
Reference
Schmid et al. (2007). International Society for Analytical Cytology biosafety standard for
sorting of unfixed cells. Cytometry A 71, 414–437.