QX100 Droplet Generator Instruction Manual

QX100 Droplet Generator Instruction Manual

| 3

Chapter 1 QX100 Droplet Generator

Chapter 1 QX100 Droplet Generator

Table 1.2. Additional materials required for droplet generation.

Component

Recommended

Catalog #

PCR supermix (required)

ddPCR supermix for probes

186-3010, 186-3026,

186-3027, 186-3028

One-Step RT-ddPCR kit for probes

186-3021, 186-3022

Control

Buffer control kit

186-3052

Droplet generator oil

Droplet generator oil (10 x 7 ml)

186-3005

Pipets

20 μl pipet for sample loading

Rainin L-20

50 μl pipet for droplet transfer

Rainin L-50, L8-50

8-channel, 200 μl pipet for oil

Rainin L8-200

Pipet tips

Filtered

Rainin GP-L10F, GP-L200F

96-well PCR plates

twin.tec semi-skirted 96-well plate

Eppendorf 951020362

Reagent trough

For droplet generator oil

Thermo 95128095

Foil plate seals

Easy Pierce foil plate seals

Thermo AB-0757

Plate sealer

Eppendorf 951023078

8-cap strips

Any

1.3 Installation and General Operation

Connect the QX100 droplet generator to a power source using only the power cord provided. Ensure the
ground is reliably connected before plugging in the instrument

Leave 10" (5 cm) space clear behind and 5" (2.5 cm) clear to the right and left for proper ventilation

Power on the droplet generator by plugging it in. The status indicator turns solid green to indicate power is on

Open and close the instrument by pressing the button on top of the green lid

U.S. Standard power cord set with grounded plug
(Type 5-15P) and C5 connector (10 A/125 V)

Power supply to 5 mm DC power jack inlet