beautypg.com

3 preparation for pcr – Bio-Rad QX100™ Droplet Digital™ PCR System User Manual

Page 16

background image

10 |

Chapter 2 Droplet Generation

QX100 Droplet Generator Instruction Manual

2.3 Preparation for PCR

1. Pipet 40 μl of the contents of the top wells (the droplets) into a single column of a 96-well plate.

Aspirating droplets from the DG8 cartridge.

Dispensing droplets into a 96-well PCR plate.

Use the following pipetting techniques to avoid shearing or coalescing the droplets:

To aspirate droplets from the DG8 cartridge:

Use an 8-channel manual P-50 Pipetman with 200 µl tips (not wide- or narrow-bore); do not use a P-20
or P-1000 Pipetman

Place the cartridge holder on a flat surface and position the pipet tips in each of the 8 top wells at
a ~30–45° angle, vertical into the junction where the side wall meets the bottom of the well. Do not
position the pipet tip in a vertical orientation (90°) or against any flat surface of the well; do not allow the
tips to be flat against the bottoms of the wells

Slowly draw 40 µl droplets into the pipet tip (should take ~5 sec, and ~5 µl air is expected); do not
aspirate >40 µl, as this causes air to percolate through the droplets

Pipet slowly. Apply a stable resistive force to the plunger to draw and aspirate droplets smoothly into
and out of pipet tips

To dispense droplets into the 96-well plate, position the pipet tip along the side of the well — near, but
not at, the bottom of the well — and slowly dispense the droplets (~5 sec).

Cover the plate (for example, with 8-cap strips or the lid from a pipet tip box) as you work to prevent
evaporation and contamination with particulates.