Chapter 4 — Error Messages and Replacing Fuses

Error Message

Explanation

Platen overheated.
Restart instrument.

If issue continues, contact

Bio-Rad.

The platen overheated. Restart the instrument.

As a safety precaution, the heater has been turned off.

Manually set the desired temperature and the platen

will begin to heat.

System did not reach

temperature.
Restart instrument.

If issue continues, contact

Bio-Rad.

The platen failed to reach target temperature in the

allotted time. As a safety precaution, the heater has

been turned off. Restart the instrument.

System overheated.
Restart instrument.

If issue continues, contact

Bio-Rad.

The internal temperature of the PX1 has exceeded

a safe limit. Power to the heater is shut down. Allow

instrument to cool and then restart.
Note: the temperature has been set to OFF. Manually

set the desired temperature and the platen will begin

to heat.

Push drawer in.
If issue continues, restart

instrument.

Drawer may be overextended. Manually push the

drawer in and touch OK. If error persists, restart the

instrument.

Platen not initialized.
Restart instrument.

If issue continues, contact

Bio-Rad.

The platen did not raise properly. Restart the

instrument to initialize the platen.

Platen blockage.
Remove object.

If issue continues, restart

instrument.

Platen failed to raise or lower because of a blockage.

Carefully remove blockage and touch OK. Ensure

the plate was placed on the correct side of the plate

support block (for example, 384-well plates placed

only on the flat side of the block and 96-well plates

placed only on the side with 96 recesses).