Home Screen Overview

The home screen (Figure 6) displays important heat sealing information.

Basic Sealing Instructions

1. Remove the plate support block and

sealing frame when the instrument is turned
on (Figure 7). Touch the

button to open

the drawer and ensure that both the plate

support block and sealing frame are removed

from the instrument. Once removed, touch OK.

The instrument will automatically begin heating

to the last set temperature.

WARNING! Burn Hazard! Do not leave the plate support block or sealing frame

in the heated instrument. They may become hot and pose a hazard to the user

and experimental samples.

2. Modify seal temperature and time

n

On the home screen touch Set Temperature

Enter desired seal temperature between 100 and 190

o

C. Touch OK

– The platen automatically begins to heat to the new set temperature. If the set

temperature is lower than the current platen temperature, the heater will remain

off and the platen will cool passively

– The Off button on the numeric keypad will turn the heater off

n

On the home screen touch Set Time

Enter desired sealing time between 0.5 and 10 seconds. Touch OK

Note: Suggested sealing parameters are provided in Chapter 5 and in Bio-Rad heat

seal packaging. Refer to the Optimizing Heat Sealing section for instructions on

optimizing these parameters.

Using the PX1

PCR Plate Sealer

2

Fig. 6. PX1 home screen.

Fig. 7. Remove support block.

Seal a plate

Open and

close drawer

Saved protocols

Instrument status

Current platen
temperature

Loaded protocol
name

Program automatic
heater shutdown

Using the PX1 PCR Plate Sealer