LaMotte AM-21 Water Pollution Detection Outfit User Manual
Page 49

Overhead Projection Demonstration
Quantity
Contents
Code
60 mL
*TDS Reagent A
4802-H
60 mL
*TDS Reagent B
4803-H
2
Resin Column
1079
15 mL
Methyl Orange Indicator with Halidex
2299-E
1
Test Tube, 5-10-15 mL, glass, w/cap
0778
60 mL
Deionized Water
5115PT-H
2
Direct Reading Titrators, 0-1000 Range
0384
1
Pipet, 1.0 mL
0354
1
Demonstration Stage, three cell
1039
1
Spatula
0691
*WARNING: Reagents marked with a * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD
or www.lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
Read the LaMotte Direct Reading Titrator Manual before proceeding.
Procedure
1.
Place the demonstration stage (0139) on the overhead projector and
turn the projector light on.
2.
Fill the test tube (0778) to the 10 mL line with sample water and transfer
to a cell on the demonstration stage (0139).
3.
Add three drops of Methyl Orange Indicator with Halidex (2299) to the
cell and stir with the clean spatula.
4.
Fill a Direct Reading Titrator (0384) with *TDS Reagent A (4802), add
one drop at a time, while stirring until the yellow color changes to pink.
Read the test result directly from the scale where the large ring on the
Titrator meets the Titrator barrel. This is result “A.” Rinse the titraton
tube.
5.
Suspend the resin column (1079) in the second test tube (0778).