Field Test Method

Quantity

Contents

Code

60 mL

*Mixed Acid Reagent

*V-6278-H

5g

*Nitrate Reducing Reagent

*V-6279-C

1

Spoon, 0.1g

0699

2

Test Tube, 2.5 and 5 mL

0820

1

Nitrate-N Comparator, 2 and 10 ppm

7493

1

Pipet

0352

1

Dispenser Cap

0692

WARNING: Reagents marked with a * are considered to be potential health hazards.

To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD

or www.lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
NOTE: Place dispenser cap (0692) on *Mixed Acid Reagent (V-6278-H). Save this cap

for refill reagents.

Procedure

1.

Fill a test tube (0820) to the 2.5 mL line with the sample water.

2.

Add *Mixed Acid (V-6278) until the tube is filled to the 5.0 mL line.
Cap and mix. Wait two minutes.

3.

Use the 0.1g spoon (0699) to add one level measure of *Nitrate
Reducing Reagent (V-6279) to the mixture in the test tube. Invert the
test tube 50-60 times in one minute. Wait 10 minutes.

4.

A very light pink color indicates a trace of Nitrate Nitrogen present in
the sample. High concentrations of Nitrates will produce a deep magenta
color. Insert the tube into the Nitrate-N Comparator (7493). Match
sample to a color standard.

34