LaMotte AM-21 Water Pollution Detection Outfit User Manual
Page 34
Field Test Method
Quantity
Contents
Code
60 mL
*Mixed Acid Reagent
*V-6278-H
5g
*Nitrate Reducing Reagent
*V-6279-C
1
Spoon, 0.1g
0699
2
Test Tube, 2.5 and 5 mL
0820
1
Nitrate-N Comparator, 2 and 10 ppm
7493
1
Pipet
0352
1
Dispenser Cap
0692
WARNING: Reagents marked with a * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD
or www.lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
NOTE: Place dispenser cap (0692) on *Mixed Acid Reagent (V-6278-H). Save this cap
for refill reagents.
Procedure
1.
Fill a test tube (0820) to the 2.5 mL line with the sample water.
2.
Add *Mixed Acid (V-6278) until the tube is filled to the 5.0 mL line.
Cap and mix. Wait two minutes.
3.
Use the 0.1g spoon (0699) to add one level measure of *Nitrate
Reducing Reagent (V-6279) to the mixture in the test tube. Invert the
test tube 50-60 times in one minute. Wait 10 minutes.
4.
A very light pink color indicates a trace of Nitrate Nitrogen present in
the sample. High concentrations of Nitrates will produce a deep magenta
color. Insert the tube into the Nitrate-N Comparator (7493). Match
sample to a color standard.
34