3.3 Installation diagram

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

850 Professional IC – AnCat – MSM-HC – MCS

3

Detector inlet capillary

4

Conductivity detector
For the cation channel.

5

Detector outlet capillary

6

Column inlet capillary
For the anion channel. Connected to the
injection valve and threaded into the capil-
lary recesses of the column thermostat.

7

Connection capillary
Capillary of the MSM-HC labeled with Elu-
ent
.

8

Connection capillary
Capillary of the MSM-HC labeled with
Detector.

9

Detector inlet capillary

10 Conductivity detector

For the anion channel

11 Detector outlet capillary

12 Sample aspirating capillary 6.1803.040

For the anion channel. Connected to the
lower injection valve. The other end can be
connected either to the sample degasser or
guided directly out of the instrument.

13 PTFE connection capillary

Section of the 6.1803.040 PTFE capillary.
Connects the sample degasser with the
Sample Processor (required only if the sam-
ple degasser is used).

14 Sample outlet capillary

For the cation channel. Connected to the
lower injection valve.

15 Sample aspirating capillary 6.1803.040

For the anion channel. Connected to the
upper injection valve. The other end can be
connected either to the sample degasser or
guided directly out of the instrument.

16 PTFE connection capillary

Section of the 6.1803.040 PTFE capillary.
Connects the sample degasser with the
Sample Processor (required only if the sam-
ple degasser is used).

17 Sample outlet capillary 6.1803.040

For the anion channel. Connected to the
upper injection valve.

18 Regeneration solution aspirating capil-

lary 6.1803.020
Connection to the regeneration solution
bottle.

19 Pump tubing 6.1826.320

Pump tubing with yellow-orange stoppers
for conveying the regeneration solution.

20 Connection capillary for the regenera-

tion solution
Capillary of the MSM-HC labeled with
H2SO4.

21 Regeneration solution outlet capillary

Capillary of the MSM-HC labeled with
Waste.

22 Rinsing solution aspirating capillary

6.1803.020
Connection to the bottle with rinsing solu-
tion.

23 Pump tubing 6.1826.320

Pump tubing with yellow-orange stoppers
for conveying the rinsing solution.

24 Connection capillary for the rinsing

solution
Capillary of the MSM-HC labeled with H20.