3 Installation
850 Professional IC – AnCat – MSM-HC – MCS
■■■■■■■■
15
8 Initial start-up
■
Switch on the PC and start MagIC Net.
■
Switch on the instrument.
■
Deaerate the high pressure pump (see Chapter 3.10.2, page
40).
■
Set contact pressure of the peristaltic pump (see "Set flow rate",
page 59).
■
Rinse instrument without column.
9 Installing guard and separation column
■
Remove the 6.2744.040 coupling between the column inlet capil-
lary and the eluent inlet capillary of the MSM-HC.
■
Connect guard column (optional) (see Chapter 3.21, page 67)
– Fasten the guard column to the end of the column inlet
capillary according to the specifications contained in the
leaflet accompanying the guard column.
– Rinse guard column.
■
Connecting the separation column for the anion channel (see
Chapter 3.22, page 69)
– Fasten the inlet of the separation column either to the end
of the column inlet capillary or to the guard column (if
used) according to the specifications contained in the leaf-
let accompanying the column.
– Fasten the MSM-HC capillary labeled with Eluent with a
6.2744.070 PEEK pressure screw at the output of the sepa-
ration column.
■
Connecting the separation column for the cation channel (see
Chapter 3.22, page 69)
– Fasten the inlet of the separation column either to the end
of the column inlet capillary or to the guard column (if
used) according to the specifications contained in the leaf-
let accompanying the column.
– Fasten the detector inlet capillary using a 6.2744.070 PEEK
pressure screw to the output of the separation column.
■
Hang the separation column for the anion channel with its chip in
the column holder A of the instrument.
■
Hang the separation column for the cation channel with its chip in
the column holder B of the instrument.