beautypg.com

Metrohm 850 Professional IC AnCat MSM-HC MCS User Manual

Page 23

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC – AnCat – MSM-HC – MCS

■■■■■■■■

13

Place the detectors in the instrument and connect them (see
Chapter 3.6.2, page 24)
.

Connect the leak sensor (see Chapter 3.6.4, page 26).

Connect drainage tubings (see Chapter 3.6.5, page 27).

3 Connecting the eluent path

Equip an eluent aspiration tubing for each of the two high pres-
sure pumps and connect with the eluent bottle (see Chapter
3.8.1, page 32)
.

Instead of the columns, connect a 6.2744.040 coupling with a
6.2744.010 pressure screw at the end of each pre-installed col-
umn inlet capillary.

Connecting the cation channel:
Connect the detector inlet capillary of the detector for the cation
channel with a 6.2744.010 PEEK pressure screw to the
6.2744.040 coupling of the column inlet capillary of the cation
channel (connected to the lower injection valve).

Connecting the anion channel

Connect the MSM-HC (see Chapter 3.16.2, page 52)

capillary labeled with Eluent with a 6.2744.010 PEEK pres-
sure screw to the 6.2744.040 coupling of the column inlet
capillary of the anion channel (connected to the upper
injection valve).
Use a 6.2744.090 long pressure screw to connect the capil-
lary of the MSM-HC (see Chapter 3.16.2, page 52)
labeled with Detector to the input of the MCS (see Chapter
3.18.2, page 60)
.

– Connect the detector inlet capillary of the detector for the

anion channel with a 6.2744.090 long pressure screw to
the output of the MCS (see Chapter 3.18.2, page 60).

4 Installing the sample path

Note

The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.

Connect the 6.1803.040 sample aspirating capillary connected at
the sample input of the injection valve to the output of the sam-
ple degasser with the aid of a 6.2744.090 long pressure screw
(see Chapter 3.13, page 45).