6 suppressor drive, 1 inserting the rotors, Suppressor drive – Metrohm 872 Extension Module Suppression – MCS User Manual
Page 29: Inserting the rotors, Figure 8, Rotors and adapter
![background image](/manuals/587098/29/background.png)
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
872 Extension Module Suppression – MCS
■■■■■■■■
21
3.6
Suppressor drive
The suppressor housing of the 872 Extension Module Suppression – MCS
can hold various rotors. Large rotors like the SPM rotors A (6.2835.000)
and the MSM-HC rotor A (6.2842.000) can be inserted directly.
Small rotors like the MSM rotor A (6.2832.000) and the MSM-LC rotor A
(6.2844.000) must first be fitted into the adapter (6.2842.010) which can
then be inserted into the suppressor housing.
For all rotors the large connecting piece (6.2835.010) is used.
6.2835.000
6.2842.000
6.2842.010
6.2844.000
6.2842.010
6.2832.000
Figure 8
Rotors and adapter
Note
The instruments are supplied without rotor and without adapter.
The suitable rotor and the adapter, if required, must be ordered sepa-
rately.
3.6.1
Inserting the rotors
Large rotors can be directly inserted into the rotor housing.
Caution
An incorrectly inserted rotor can be destroyed during start-up.
Therefore, follow the following instructions exactly.