beautypg.com

8 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 886 Prof.Thermostat/Reactor User Manual

Page 49

background image

■■■■■■■■■■■■■■■■■■■■■■

8 Troubleshooting

886 Professional Thermostat (2.886.0110) / 886 Professional Reactor (2.896.0210)

■■■■■■■■

41

8 Troubleshooting

8.1

Problems and their solutions

Problem

Cause

Remedy

Marked drop in
pressure

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.4, page 17).

Drift of the baseline

Thermal equilibrium not yet
attained.

Condition instrument with the column heater
switched on.

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.4, page 17).

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Capillary connections –
dead volume in the system.

Check connections (see Chapter 3.4, page 17)
(use PEEK capillaries with an internal diameter
of 0.25 mm between the injection valve and
detector).