3.2 Initial installation
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
850 Professional IC Anion MSM-HC HP-Gradient
■
Connect the 6.1803.020 aspirating capillary for rinsing solutions
with a 6.2744.034 tubing olive and a 6.2744.070 short pressure
screw to the aspiration end of the second 6.1826.320 pump tub-
ing for the rinsing solution.
■
Place the second pump tubing in the second tubing cartridge.
■
Place both tubing cartridges into the peristaltic pump.
6 Connecting the MSM-HC
■
Connect the capillary labeled with regenerant with a 6.2744.180
pump tubing connection and a 6.2744.070 short pressure screw
to the peristaltic pump on the outlet end of the pump tubing for
the regeneration solution.
■
Connect the capillary labeled with Rinsing solution with a
6.2744.180 pump tubing connection and a 6.2744.070 short
pressure screw to the peristaltic pump on the outlet end of the
pump tubing for the rinsing solution.
■
Guide the two capillaries labeled with Waste reg. and Waste rins.
into a waste container and fasten them there.
7 Connecting the instrument
■
Connect the instrument to the PC with a 6.2151.020 USB cable
(see Chapter 3.18.1, page 55).
■
Connect instrument to mains supply (see Chapter 3.18, page
55).
8 Initial start-up
■
Switch on the PC and start MagIC Net.
■
Switch on the instrument.
■
Deaerate the high pressure pump (see Chapter 3.10.2, page
36).
■
Set contact pressure of the peristaltic pump (see "Set flow rate",
page 54).
■
Rinse instrument without column.
9 Installing guard and separation column
■
Remove the 6.2744.040 coupling between the column inlet capil-
lary and the MSM-HC capillary labeled with in.