8.8 Remote control via an RS-232 connection

■■■■■■■■■■■■■■■■■■■■■■

62

■■■■■■■■

876 Dosimat plus

Command

Function

Comment

$Q(variable)

Request variable value

List of variables, see below

Variable

Comment

VOLUME

Dosed volume

TITER

Titer of selected solution

CONC

Concentration of selected solution

C00

Sample size (DOS only)

RESULT

Result (DOS only)

RATE

Dosing rate (XDOS only)

TIME

Dosing time (XDOS only)

The values of the variables are only available after the end of a determina-
tion (in the status 'ready').

Answer of the instru-
ment

Comment

OK

Command executed

E1

Method not found

E2

Invalid variable

E3

Invalid command