4 calculation, Calculation, Figure 17 – Metrohm 876 Dosimat plus User Manual
Page 49: Dosing ramp, two examples

■■■■■■■■■■■■■■■■■■■■■■
6 Parameters and dosing modes
876 Dosimat plus
■■■■■■■■
41
Dosing ramp
The dosing ramp is a delay in the dosing rate at the time of the start of a
dosing step. It is advantageous primarily for manual titrations with indica-
tor, when small volume steps are to be dosed before the equivalence
point. The selectable delay time determines after which amount of time (in
seconds) the defined dosing rate is to be reached.
Input range
0 ... 10 s
Default value
0 s
2
4
6
8
10
s
30
60
90
mL/min
Figure 17
Dosing ramp, two examples
6.1.4
Calculation
Menu
▶ Parameters ▶ Calculation
The calculation formula is predefined and cannot be edited.
Calculation formula
(Volume - Blank) × Titer × Conc. × Fac-
tor / (Sample size × Divisor)
where:
Volume
Dosed volume
Blank
Blank value
Titer
Titer of the solution used
Conc.
Concentration of the solution used
Factor
Factor
Sample size
Sample size. The sample size and the associated unit
can be entered under Menu
▶ Sample data. If a
balance is connected, then the sample size (with
unit) can also be accepted directly from the balance.
Divisor
Divisor