Application Bulletin 356
Installation instruction for «VoltIC pro I»
Page 13 of 18
Metrosep A Supp 5 - 150/4.0 to the analysis anion, assign
the column Metrosep C 4 - 100/4.0 to the analysis cation,
assign the two eluents to their respective IC pump, match
the two suppressor solutions to the IC peristaltic pump, and
define the rack of the connected sample processor. Adjust
the temperature of the column oven to 30°C. Save the
method.
In order to start the equilibration, go to the window work
place, load your method, and press «Start HW». Before
inserting the columns, flush the system for about 10 minutes
and get rid of air bubbles (by using the purge valve and
syringe), then insert and rinse the precolumns for 10
minutes, and only afterwards put the analytical columns into
the flow (see also 850 manual).
3.4
Installation of 797 VA Computrace software
Install the software. All the standard directories proposed by
the program should be accepted. Restart windows.
Connected USB devices are automatically recognized when
VA Computrace software is started.
3.5
Configuration of 797 VA Computrace software
Make the following general settings in the main window
under «Settings» «General Settings»
s” and click «OK»:
Load method «Det of Zn Cd Pb Cu_IC_VA.mth», modify or
create a new method.
Click on «Dosinos» in the «Working Method Specifications»
window and make following settings: