Metrohm VoltIC pro I User Manual
Page 12

Application Bulletin 356
Installation instruction for «VoltIC pro I»
Page 12 of 18
For the data transfer from the VA Computrace software to
MagIC Net
™
an RS 232 device must be added to the device
list as follows:
Go to the Configuration window in MagIC Net
™
and select
«New» from the «Edit» button in the Device window. Select
the «RS 232 Device» from the «Miscellaneous» list:
Rename the device «797 RS232 Printer». Enter «1» in the
field «serial number».
Make the following settings under «RS 232»:
(Attention: the second newly generated COM port by the
null modem cable has to be selected as «COM Port», e.g.
COM 15 !)
Enter «stopp» in the field «Terminator for receive».
In the window method, import the method for VoltIC pro I
from the installation CD: Go to File
→ method manager and
choose your method group. Afterwards, click on edit
→
import and choose the pathway on the installation CD
(MagICnet\examples\methods\VoltICpro.imet).
The method carries out a parallel determination of cations
and
anions.
Immediately
after
starting
the
anion
determination, the cation determination (of the same
sample) is initialized. Samples could be taken from the
same vial, or from two vials with one of them containing the
acidified sample solution for cation determination. Both
measurements are recorded in the same data file. In order
to measure samples from two different vial please select
method «VoltICpro_mod.imet» from the same folder.
In the window method, open the VoltICpro method and
assign the following equipment: Assign the column