beautypg.com

Metrohm NIRS XDS MultiVial Analyzer User Manual

Page 77

background image

▪▪▪▪▪▪▪

75

This view shows the fill width of the tabulated results, and expands the spectrum for better viewing.

When Performance Test is finished, a message box is displayed
to indicate test completion and status. All measured values are
compared with acceptance criteria stored in Vision. If all results
meet acceptance criteria, the test is successful and this dialog
box is displayed.

Click “Print Report” for a hard copy of results. Before clicking
“Close Report”, the user is directed to the tabular display.

To view the tabular display of results, place cursor over the tabular display and double-click twice.
Vision enlarges the tabular portion of the screen. Now click on the OpQual tab, near the bottom of
the screen. Noise summary results are shown.