beautypg.com

Metrohm NIRS XDS MultiVial Analyzer User Manual

Page 67

background image

▪▪▪▪▪▪▪

65

If Vision displays the Instrument
Configuration screen again, click “OK to
accept the instrument identification.

Vision briefly displays a dialog box which indicates that the instrument reference is being placed
into position. In reflectance, this is automatic. No user action is required.

The status bar indicates scan progress.

Vision requests the “Standard File” for the
Certified Wavelength Standard.

Place the mini-CD into the CD drive, and select
the standard file. Click “Open”.

Vision requests that the Certified 80%
Reflectance Reference Standard be placed in
the sampling area.

Place the standard as shown, with the
identifying label parallel to the long axis of
the tray.

Close the cover, and click on “OK”.