At the bottom of the Vial Dialog screen is a
checkbox marked “Use Order”.

If this box is checked, the MultiVial Analyzer will
start with the sample marked “1” in the
“Order” column, and proceed numerically
upward through the order designated in the
“Order” column.

If this box is left unchecked, the MultiVial
Analyzer will proceed from vial position 1, to 2,
to 3, in numerical order of the physical vial
position in the tray. Vision will ignore the
“Order” column.

If the user plans to enable “Use Order” it is
important to set the order correctly.

Fortunately there is a method to set order easily, without losing Sample ID information.

In the example above, the order is out of sequence for samples, 7, 9, and 16. If “Use Order” is left
unchecked, Vision proceeds in order of Sample Number. No action is needed.

However, if the user wishes to set the sequence, the procedure is as follows:

1.

Click on each vial position to be changed.

2.

Re-click, and that vial’s “order” will be sent to the end of the sequence.

3.

Because this may be difficult to achieve by setting one vial at a time, click on all vial positions,

to turn them white. Then click on each one, in the order desired. This will set the “Order”