Metrohm NIRS XDS MultiVial Analyzer User Manual

Page 69

background image

▪▪▪▪▪▪▪

67

A final spectrum (green
when plotted on-screen) is
plotted to verify that the
corrected spectrum
produces the same results
as the Certified 80%
Reflectance Standard.

Click “Close Report” to
continue. The correction is
automatically downloaded,
and is saved in the
Diagnostic Data Base.

The correction will be
applied in real time to all
spectra taken – in the “Use
Iris” mode --with a DCM
where “Reference
Standardization” is
checked.

When finished, remove the Iris Adapter, and return to the “Use Tray” mode.

NOTE: Perform Reference Standardization immediately after changing the spot
size, to have the proper reference correction loaded in the instrument.

In the event that Vision does not find a
Reference Standardization where it is
needed, it will prompt for a new one
upon connection as shown. When a
new Reference Standardization is
performed, Vision asks for a new
Instrument Calibration as well.