3 installation, 1 about this chapter, 2 initial installation – Metrohm 882 Compact IC plus – Cation User Manual
Page 17: About this chapter, Initial installation

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
882 Compact IC plus – Cation
■■■■■■■■
9
3 Installation
3.1
About this chapter
The Installation chapter contains:
■
this overview.
■
a brief set of instructions for the initial installation of the 882 Compact
IC plus – Cation. At each step you will find cross-references to more
detailed installation instructions for individual components, should you
require such aids.
■
an installation diagram (see Figure 3, page 12), showing a com-
pletely installed 882 Compact IC plus – Cation.
■
several chapters with detailed installation instructions for all compo-
nents, including those that are already installed at the time the instru-
ment is delivered.
3.2
Initial installation
Note
A number of the capillaries is already connected at the time the instru-
ment is delivered.
Installing 882 Compact IC plus – Cation
Install the instrument as follows:
1 Setting up the instrument
■
Place the instrument in a location of the laboratory which is suita-
ble for operation and free of vibrations.
The installation site must be protected against corrosive atmos-
phere and contamination by chemicals. If possible, avoid excessive
temperature fluctuations and direct sunlight.
2 Installations on the rear of the instrument
■
Place the detector in the instrument and connect it (see manual
of the detector).
■
Remove transport locking screws (see Chapter 3.6.1, page 16).