beautypg.com

Metrohm NIRS Analyzer PRO User Manual

Page 56

background image

54

▪▪▪▪▪▪▪

5.

When the test is finished, the

split-screen display will
appear as shown. (The test
normally takes 9-10 minutes
in this configuration.)

Double-click in the lower
right quadrant to see the
tabulated results.

6. The tabulated results show these
items:

Scan:

10 is the default value.

EOC:

Errors on communication – this will be “0” nearly all the time.

P-P:

Peak to Peak noise

Minimum:

Magnitude of lowest-going noise peak

Wavelength: Location of negative peak

Maximum: Magnitude of highest-going noise peak

Bias:

Baseline offset of noise scans

RMS:

Root-mean-square value of P-P

Gain:

Not used on this instrument