Page 13

▪▪▪▪▪▪▪
11
3
Specifications
Overall Analyzer PRO instrument specifications are as follows:
Item
Specification
Ambient Temperature
23 to 104° F (-5 to 40° C)
With Air Cooling: 23 to 149° F (-5 to 65° C)
Ambient Humidity
10-90% Relative Humidity, non-condensing
Electrical Supply
100-240 VAC, 50-60 Hz, 2.0 A, 150 W
Protection
IP69k according to IEC 60529 and DIN 40050 part 9, NT ELEC 023
Dimensions
Width: 17.0” (42 cm)
Height: 17.0” (42 cm)
Depth: 5.0” (13 cm)
Weight
33.0 pounds (15 Kg)
Pressurized Air Cooling
(Ambient Temperature 45-
65 Degrees C)
Cooling air flow rate minimum 5 Liters/min
>99.9% water free
>99.9% free of oil and fine particles down to 0.3 micrometers
Pressurized Air – Spoon
Probe
Purge air: 3-5 bar
Network Connections
Local Area Network (LAN) – Ethernet, RJ-45 connection
Sample Temperature,
Reflectance Window
302 degrees F (150° C)
Sample Temperature,
Spoon Probe
248 degrees F (120° C)
Computer Requirements
See Vision computer requirements
Wavelength Range
1100-1650 nm
Analysis Time:
5 – 50 ms / integration time; depends upon sample. Typical result
time: 3-15 sec.
Wavelength Accuracy
0.5 nm
Wavelength Precision
< 0.02 nm
Wavelength Stability
< 0.01 nm/°C
Specifications for each type of sample interface follow: