mains cable is replaced, the same type of cross-section of cable must be used to assure a good seal.

The cables are carefully routed to prevent interference with operation of the Analyzer PRO.

When replacing the temporary AC mains cable, follow this procedure:

1.

Remove power from the instrument.

Both poles must be broken.
Use safety lock-out procedures if
applicable.

2.

Unplug or disconnect existing,

temporary AC mains cable.

3.

Remove the 32 screws holding the cover

in place on the instrument. Use a 7 mm
socket or nut driver.

NOTE: If using a powered removal tool,
set the torque on a low setting, to avoid
stressing the screws. Start at a low
torque, then increase enough to remove
the screws easily. Use a similar torque
when re-installing the screws. Avoid
cross-threading.