beautypg.com

10 servicing the pulsation absorber, 11 injection valve, 12 suppressor – Metrohm 930 Compact IC Flex Oven/ChS/PP/Deg User Manual

Page 89: 1 notes for operating the suppressor, Notes for operating the suppressor

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

930 Compact IC Flex Oven/ChS/PP/Deg (2.930.2360)

■■■■■■■■

81

5.10

Servicing the pulsation absorber

CAUTION

The pulsation absorber is maintenance-free and must not be opened.

5.11

Injection valve

Maintenance on the injection valve is best performed by specialist person-
nel from Metrohm during annual service.

5.12

Suppressor

5.12.1

Notes for operating the suppressor

NOTE

The suppressor units must never be regenerated in the same flow direc-
tion the eluent is pumped. Therefore, always mount the input and out-
put capillaries according to the diagram outlined in Chapter Connecting
the suppressor, page 32
.

The suppressor consists of three suppressor units, which are used in rota-
tion for suppression, regenerated with regeneration solution, and rinsed
with ultrapure water. In order to record every new chromatogram under
comparable conditions, you should normally work with a freshly regener-
ated suppressor.

CAUTION

The suppressor must never be switched over if liquid is not flowing
through it, since otherwise it may become jammed. If the suppressor is
in a dry state, it must be rinsed for at least five minutes before it may be
switched over.