7 Technical specifications

930 Compact IC Flex Oven/ChS/PP/Deg (2.930.2360)

■■■■■■■■

103

7.13

Sample degasser

Material

Fluoropolymer

Resistance to sol-
vents

No restriction (except PFC)

Time to establish
vacuum

< 60 s

7.14

Power connection

Required voltage

100 - 240 V ± 10% (autosensing)

Required fre-
quency

50 - 60 Hz ± 3 Hz (autosensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

Internal fuse 3.15 A

7.15

Interfaces

USB

Input

1 USB upstream, type B (for connection to the PC)

Output

2 USB downstream, type A

MSB

2 MSB Mini DIN 8-pin (female) (for Dosino, stirrer, remote lines, etc.)

Detector

1 15-pin high density DSUB (female)

Column recogni-
tion

for an intelligent column

Leak sensor

1 jack plug

Further connec-
tions

Auxiliary

1 15-pin DSUB (female)

Service