Metrohm 930 Compact IC Flex Oven/ChS/PP/Deg User Manual
Page 104
6.1
■■■■■■■■■■■■■■■■■■■■■■
96
■■■■■■■■
930 Compact IC Flex Oven/ChS/PP/Deg (2.930.2360)
Problem
Cause
Remedy
Guard column – Blocked.
Replace the guard column (see Chapter 3.19,
page 49).
Separation column –
Blocked.
■
Regenerate the separation column (see
Chapter 5.16.4, page 93).
■
Replace the separation column (see "Con-
necting the separation column", page 53).
Note: Samples should always be microfiltered .
Injection valve – Valve
blocked.
Have the valve cleaned (by a Metrohm service
technician).
The retention times
in the chromato-
gram have changed
unexpectedly.
Separation column –
Diminished separating effi-
ciency.
■
Regenerate the separation column (see
Chapter 5.16.4, page 93).
■
Replace the separation column (see "Con-
necting the separation column", page 53).
The eluent contains gas
bubbles.
■
Ensure that the connectors for the eluent
degasser are firmly connected .
■
Deaerate the high-pressure pump (see Fig-
ure , page 48).
High-pressure pump –
Defective.
Request Metrohm Service.
Peak areas are
lower than
expected.
Sample – There is a leak in
the sample path.
Find and fix the leak in the sample path.
Sample – The sample path
is blocked.
Find and fix the blockage in the sample path.
Sample – The sample loop
is not (completely) filled.
Adjust the transfer time for the sample (see
"Determining the transfer time", page 91).
Sample – There are gas
bubbles in the sample.
Use the sample degasser .
The peristaltic pump
is pumping too little.
Peristaltic pump – Contact
pressure too weak.
Correctly set the contact pressure .
Peristaltic pump – Filter
blocked.
Replace the filter (see "Replacing the filter",
page 90).
Peristaltic pump – Pump
tubing defective.
Replace the pump tubing .