3 silo memory, 1 display silo memory – Metrohm TiNet 2.5 User Manual
Page 43

TiNet 2.5 Compliant
Titrations • 41
5.3 Silo
memory
5.3.1
Display silo memory
Every parallel titration has its own silo memory, which is shown on the
Silo
#
tab in the main window. The silo memory is used on one side for the
display of the finished determinations and their results, and on the other
side for entering sample data "for future use" (e.g. for sample series with a
sample changer).
The silo memory contains the
Status
column and the columns defined un-
der
View sample data
(see section 5.2.2) and
View results
(see section
5.3.4).
Status
Display of the determination status:
ready
The sample data are used for the next
determinations.
running
The determination is running.
finished
The determination is finished, results have
been
entered.
aborted
The determination has been aborted.
For automatic processing of sample data from the silo memory for series
determination, the
Silo
option on the
Sample data
tab must be enabled and
the number of automatic starts must be entered in the
Autostart
field. In
this case, the sample data are transferred from the silo memory at each
new method start. The weighing data from a connected balance are writ-
ten to the silo memory. The silo memory operates on the FIFO (first in, first
out) principle: The lines are entered and processed in ascending order.
If the silo memory is enabled, the sample data can be edited using con-
text-sensitive menu items (see section 5.3.2). The content of the silo mem-