Metrohm TiNet 2.5 User Manual
Page 11

TiNet 2.5 Compliant
Overview • 9
6. Select the option
TiNet 2.5 Compliant (with 21 CFR 11 features)
and click
on
Next>
.
7. Select the installation directory and click on
Next>
. The "TiNet 2.5
Compliant" program will be installed.
8. Click
on
Finish>
to complete the installation.
9. Open the TiNet 2.5 Compliant program. The
Login
window appears
(see section 7.1.1).
10. Enter "
Admin
" for
User name
and "
TiNet2
" for
Password
. Click on
.
11. Select
Extra/Options/Print options
to open the
Print options
window (see
section 3.4.2).
12. In the
Print options for TiNet reports
box, enable the
Special printer
op-
tion and select the
FinePrint pdfFactory Pro
printer driver.
13. Select the directory to which PDF reports will be saved. There must be
no ‘.’ (dot) in the whole path!
14. Select a
Delay
for the PDF creation. It is not recommended to select
shorter times than 45 sec. Note that this is the minimum method dura-
tion time (add a WAIT block to the method if the duration time is
shorter than this delay)!
15. Enable or disable the option for immediate printing of PDF reports.
This requires the installation of Adobe Acrobat Reader (see Step 1).
16. Click on
to close the
Print options
window and close the pro-
gram.
17. Create a new method containing a WAIT block (waiting time 60 s), a
CALC block (with a calculation formula "RS = SmplSize") and a
REPORT block (text field, "Print immediately" activated).
18. Start the method.
19. If you are working with a NTFS file system, make sure that all Windows
users have read and write privileges for all subdirectories of the pro-
gram directory, for the method directory, for all export directories de-
fined in the program and for C:\ and the two files
WIN.INI
and
TINET2.INI
in the Windows directory.